The Materials Science and Powder Diffraction Beamline (MSPD) has two endstations: one for high angular resolution powder diffraction and another for for high pressure powder diffraction using diamond anvil cells and micro powder diffraction at high energy.

im-mspd-station1.jfif

The beamline operates between 8 and 50 keV. This range adequately covers the desirable energy for almost any kind of powder diffraction experiments in transmission geometry. The high energy part makes posible total scattering experiments and high pressure diffraction using Diamond Anvil Cells. The easily tunable energy allows selecting adequate conditions for powder diffraction in transmission geometry: optimize absorption, avoid absorption edges.

Research Area

  • Materials Science
  • Mineral processing and soil sciences
  • Pharma
  • Engineering

Technical Specifications

Photon Energy Range 

 8 - 50 keV

Flux at sample

~ 4·1012 ph/s 

Energy resolution

2·10-4

Beam size at sample
collimating
focus
KB

variable:
5 x 1.5 mm2
5 x 0.7 mm2 
0.015 x 0.015 mm2

Beam divergence at sample 

variable

How to download the data

Check out section 4 to access remotely to your experimental data