ALBA Synchrotron
The Materials Science and Powder Diffraction Beamline (MSPD) has two endstations: one for high angular resolution powder diffraction and another for for high pressure powder diffraction using diamond anvil cells and micro powder diffraction at high energy.
The beamline operates between 8 and 50 keV. This range adequately covers the desirable energy for almost any kind of powder diffraction experiments in transmission geometry. The high energy part makes posible total scattering experiments and high pressure diffraction using Diamond Anvil Cells. The easily tunable energy allows selecting adequate conditions for powder diffraction in transmission geometry: optimize absorption, avoid absorption edges.
Technical Specifications
Photon Energy Range | 8 - 50 keV |
Flux at sample | ~ 4·1012 ph/s |
Energy resolution | 2·10-4 |
Beam size at sample | variable: |
Beam divergence at sample | variable |
How to download the data
Check out section 4 to access remotely to your experimental data
View moreGive us your feedback
If you have performed experiments at ALBA, ensure that all your past experimental reports and related publications are uploaded
Access the ALBA User Office Portal