Trouble 1.1: cats, collect, oav or fluoscan applications do not open.
Cause: Control system (probably the Pool) is not running.
Solution: Please contact you local contact or the floor coordinator.
Trouble 1.2: Some buttons in applications are orange instead of green or blue.
Cause: A process could not be terminated normally, but is not necessarily still faulty.
Solution: Simply try the action you wanted to execute again. Abnormal termination was most probably a transient.
Trouble 1.3: The image shown at MXCuBE remains blue.
Cause: This version is normally used during beam centering to inspect accurately the intensity of the beam.
Solution: Simply deselect the BeamView option located on the right side of the image.
Trouble 1.4: The green ellipse representing the beam and the red axes are not showed at the oav image.
Solution: Go with the mouse to the Item list located on the right side of the oav image. Just click over the Image item (do not deselect the item!) and move the item downwards using the blue arrows bellow the item list. The Image item must be bellow the items fwhm beam, vcursor and hcursor.
Trouble 1.5: The image shown at the oav widget remains black after restarting the application.
Cause: After restarting oav, the widget shows a huge magnification of the image that does not allow to see the pin.
Solution: Simply press the Esc button on the keyboard or click the central button of the mouse. If the backlight is switched on the pin will be visible.
Trouble 1.6: The legend of the buttons in the different applications changes to "door_sats".
Cause: Probably the macroserver was restarted. This is not an error message but the spock door used to interact with the control system.
Solution: Simply close the GUI and restart it again. To launch the different applications see the section Running your experiment or do it through the controlator window.
Trouble 1.7: Steps to carry out a Raster scan.
- Choose below the directory where you want to save your raster images.
- Click the button containing a grid on top of the oav image (4th from the left).
- Select in the oav image the area of interest where you want to perform the raster scan. Typically over the loop to find the crystal.
- On the right side of the oav gui you can select the H or V steps that will increase or decrease the number of "beams" included in you grid. Remember that each green ellipsoid represents the beam size.
- Press the raster button on the right side of the raster tool. The raster scan starts to work. The color code goes from red (no diffraction) to green (diffraction). The ice rings can be misunderstood and give rise to green ellipses so in this case check the images.
- De-select the Raster item from the item list once finished the raster scan in order to remove the grid from the image.
Trouble 1.8: The image freezes in MXCuBE
Cause: Probably the camera device server (BZOOM server on ibl1303) is not running.
Solution: Please contact you local contact or the floor coordinator.
Trouble 1.9: The data collection does start after pressing the Collect Queue button in MXCuBE.
Cause: After centering the sample it is needed to save the centre. This is not required for the Collect Now option.
Solution: Press the SAVE button located on the right side of the image.